application areas
the 2100plus chemical oxygen demand (codcr) water quality online analyzer is designed for online monitoring of chemical oxygen demand (codcr) in various water sources, including surface water, industrial wastewater, and municipal sewage.
detection principles
a known amount of potassium dichromate solution is added to the sample. in a strong acid medium, with silver sulfate as a catalyst, after high-temperature digestion, the increase in absorbance of trivalent chromium ions in the sample is directly proportional to the chemical oxygen demand (codcr). the absorbance is measured at a specific wavelength to quantify the codcr.
features
low reagent consumption, resulting in low operating costs.
modular design for easy operation and simple maintenance.
patented intelligent dilution with automatic range switching function.
equipped with power-off protection and automatic recovery functions.
flexible editing of analysis, calibration, quality control, and other processes.
ultra-high precision syringe pump to ensure accurate quantification of samples and reagents.
automatic and remote quality control functions, including standard solution verification and spike recovery test.
technical specifications
detection principle: rapid digestion spectrophotometry.
detection range: 0~200 / 2 000 mg/l, with additional selectable ranges.
measurement modes: periodic mode, controlled mode, hourly mode, continuous mode, manual mode.
verification modes: remote verification, manual verification, automatic verification (triggered periodically or by exceeding limits).
signal output: 4~20 ma, rs232 / rs485.
environmental requirements: 5~40 ℃, rh≤95%.
power requirements: 220±22 vac, 50±0.5 hz.
cabinet dimensions: approximately 520 mm (w) × 380 mm (d) × 800 mm (h).
instrument weight: approximately 45 kg (excluding reagents).